IFU for Medical Devices

Creating instructions for use (IFU) for medical devices and in vitro diagnostic products requires quite a lot of responsibility.

In fact, you need to combine your techcomm skills with project management, administrative and legal skills. Luckily, what needs to be included in the instructions for use for medical devices has been determined in relevant legislation.

If there is any industry in which it is important to make sure that users can use devices safely, effectively and efficiently, it is the medical industry.

And let's not forget the importance of decreasing your liability by means of adequate instructions for use.

If you want to create a medical device user manual you need to be aware of both the content to include and the process to follow.

In this guide I’m going to show you everything you need to know when creating IFU for medical devices.

I will not only show you how to create IFU for the U.S. market, but also for the European market.

And I will show you for both markets what the requirements are for medical devices as well as for in vitro medical devices.

I hope the article provides sufficient medical device instructions for use guidelines and that it gives deep insight into relevant regulation on what you have to include in your IFU, in order to meet relevant EU regulations and/or the Code of Federal Regulations (CFR) in the U.S.

Understanding and applying the requirements will help you pass Notified Body conformity assessment (in Europe) and FDA approval in the U.S.

We’ll start with general labeling requirements in the U.S. If you want to go directly to IFU guidelines for the U.S. click here; for IFU requirements of in vitro devices in the U.S. click here.

For the IFU requirements on medical devices in Europe click here and for creating the IFU for in vitro devices for Europe click here.

Or use the below table of contents to navigate through the article.

But I recommend reading it all :)

Let’s dive right in!

DO YOU WANT A USER MANUAL TEMPLATE THAT ALREADY CONTAINS THE LEGAL PARTS?

Take the shortest way to a compliant manual. We have developed user manual templates for medical devices (EU and US) that contain all legal content.

Table of Contents

PART 1 - Instructions for Use (IFU) in the U.S.

In the U.S., laws are drafted by Congress.

Once Congress has enacted a law, the appropriate federal agency may create the regulations or rules to implement the law.

The different U.S. Federal Agencies regulating products and situations are divided into Hazards/Safety/Firearms, Vehicles/Vehicle-Related Products, Health/Body, and Other.

Commonly Requested U.S. Laws and Regulations

The Health and Body group oversees agencies that cover alcohol, tobacco, food, cosmetics and medical devices.

Safety of medical devices and in vitro diagnostic products in the U.S. is regulated by the FDA.

The Federal Food, Drug and Cosmetic Act (FFDCA) is the law under which the FDA takes action against regulated products.

Based on the FFDCA, the FDA has created the CFR - Code of Federal Regulations Title 21. It contains the federal regulations for each product regulated by the FDA, including foods, pharmaceuticals, veterinary products and medical devices.

CFR - Code of Federal Regulations Title 21

What are the Requirements for IFU for Medical Devices in the U.S.?

Instructions for Use (IFU) are part of the labeling regulations. CFR labeling requirements for medical devices give requirements on what to communicate on labels, and include things as the name and address of the manufacturer, intended use, directions for use, handling instructions, storage information, and installation information. The FDA’s guidance document contains more specific information on the instructions for use.and includes information on risk/benefit information, easy navigation, checkout instructions, monitoring informnation, cleaning instructions, and maintenance information.

I will discuss everything that falls under the CFR labeling requirements first. The FDA’s guidance document, will be discussed afterwards.

What falls under labeling?

Labeling’ is defined as 'all labels and other written, printed, or graphic matter (1) upon any article or any of its containers or wrappers, or (2) accompanying such article' at any time while a device is held for sale after shipment or delivery for shipment in interstate commerce.

Booklets, brochures, posters, tags, pamphlets, circulars, instruction books, direction sheets, fillers, etc. are also considered to fall under labeling. So the term 'accompanying' indicates that labeling is more than just the physical association with the product.

website FDA

The labeling regulations for medical devices are found in several sections of Title 21 of the Code of Federal Regulations (CFR). This same overview can be found on the FDA’s website.

medical device

In general, U.S. CFR labeling requirements for medical devices are more about what to communicate on labels, than containing requirements on what to put in the instructions for use.

That’s why I will discuss everything that falls under labeling in the next sections first.

The FDA’s guidance document, which I will discuss as well in this article, contains more specific information on the instructions for use.

Apply the General Device Labeling Regulations for Medical Devices

Part 801 of the CFR - Code of Federal Regulations Title 21 contains General Device Labeling requirements.

These requirements concern:

Let’s have a closer look at all of these requirements.

Provide the name and address of the manufacturer on the label

The label of a medical device in package form must state the name and place of business of the manufacturer, packer, or distributor.

When a corporation places the medical device on the market, only the actual corporate name shall be used. This name may be followed or preceded by the name of the specific division of the corporation.

Abbreviations, such as "Inc." etc., may be used and it is allowed to omit "The".

In cases where a medical device is not made by the entity whose name is put on the label, a phrase must be added that indicates the role of the entity, such as "Manufactured for ___" or "Distributed by _____".

The full business address must be given on the label: it must include the street address, city, State, and Zip Code.

label medical device

Give a clear description of the intended uses

According to the CFR, the words intended uses refer to ‘the objective intent of the persons legally responsible for the labeling of devices.’

The IEC/IEEE 82079-1 standard describes the intended use as ‘an exhaustive range of functions or foreseen applications defined and designed by the supplier of the product.’

For example, the intended use of the Philips Ingenia MRI Scan is described as follows:

intended use

And the intended use of the Medtronic Model 5392 Pacemaker:

intended use pacemaker

The intended use is what the device is developed for and says what it should be used for or how it can be used.

The intended use is often given in the instructions for use.

But also any claims on the device or on any product labeling, in advertisements or in oral or written statements are considered to be the intended use.

So always be careful what you communicate!

It is allowed to change the intended use of a device or product.

If, for example, a distributor sells a product that is intended to be used for a different purpose, then this distributor is allowed to do so.

In this case, the distributor, or seller, must provide adequate labeling in accordance with the new intended uses.

If a manufacturer is aware of any reasonably foreseeable misuse or use other than the new intended use, he/she must provide adequate labeling for such a device which clearly indicates how the product article should be used.

For example, the following screenshot, taken from a video, clearly states that the face mask sold by this distributor is not intended for medical purposes.

Provide adequate directions for use

It is considered to be important that a layperson receives adequate directions that enables him/her to use the medical device safely and according to its intended use.

directions for use

There may be several reasons why directions are seen as inadequate.

Such as when the following information is (partly) missing or incorrect:

This example was found on the website of the FDA and represents an incorrect specification of the intended use

specification intended use

In this example a description of the intended use fails on the cover page:

intended use medical

Do not give misleading statements

A false or misleading representation with respect to another device, drug, food or cosmetic is considered to be a misleading statement.

Therefore, always make sure that the labeling of your product is authentic and does not look like other products.

name

Give label statements a prominent position

Section 502(c) of the FFDCA states that all required information should have a prominent position.

FFDCA

When there is a lack of prominence of a word, phrase or any other detail which must be displayed on the label, it may fall short of the FFDCA Act.

For example, the absence of the word, phrase, or detail in question on at least two sections of the label is considered to be a labeling defect.

The required statements must cover the entire area that is intended for placing the information.

Make sure that the label occupies adequate space on the product, and be visible enough to be clearly seen. Also, any other information shall not be more visible than the required information.

Apply a font size and type that is clearly legible and use sufficient contrast between the background and font color. Use a common, neutral design, that is not overwhelmed with other written or graphic content.

covid

While there might be exemptions due to insufficiency of label space, no exemption will be considered if the label space was used for a phrase, word, detail device, or design that is deemed unnecessary or irrelevant under the FFDCA Act.

Also, in case of insufficiency of space, no other language shall be put on the label other than the English Language.

In regions where English is not the main language of communication, such as Puerto Rico, English may be replaced by the commonly spoken language of that region and when prescribed by a medical professional.

When symbols are used, they must be followed by a very descriptive and simplified note (for better understanding), written in the official language of the region.

A symbol can come without explanatory text, when it comes from an FDA approved standard, is used according to the FDA’s specifications for use and is explained in a symbol glossary.

A symbol developed by a Standards Development Organization, such as ISO, can also be used without an explanation, when:

symbolen

Use YYYY-MM-DD to format dates

Dates included on a device’s label like expiry dates and production dates must have the approved YYYY-MM-DD formatting, with Y representing the year, M representing the month, and D representing the date.

Therefore, a product with expiry date March 26, 2022, for example, must look like this; 2014-03-26.

This is a typical guideline. These are cases of exemption though:

labels

Place a Unique Device Identifier (UDI) on the label

Labels of all medical devices must bear a so-called Unique Device Identifier (UDI).

There are a few exceptions however.

Amongst others, finished devices manufactured and labeled prior to the compliance date established by the FDA, class I medical, that are exempted by the FDA from the good manufacturing practice requirements, some individual single-use devices, devices used solely for research, teaching, or chemical analysis(and not intended for any clinical use), veterinary medical device, custom devices, investigational device and devices intended for export do not need to bear the UDI.

The UDI must contain easily readable plain text and shall make use of automatic identification and data capture (AIDC) technology: a technology to automatically identify objects, collect data about them, and enter them directly into computer systems, without human involvement.

The UDI must include a lot or batch number, a serial number, a manufacturing date, an expiration date or any other element allowing the device’s identification.

The FDA or FDA-accredited issuing agencies issue specific UDI systems. The UDI must be issued under one of these official systems.

The UDI must comply with ISO/IEC 15459-2, ISO/IEC 15459-4 and ISO/IEC 15459-6, and use only characters and numbers from the invariant character set of ISO/IEC 646.

Device data can be submitted to GUDID.

GUDID

Stand-alone software may also be regulated as a medical device.

The UDI of stand-alone software that can be downloaded (so which is not packaged) must be displayed in easily readable plain-text whenever the software is started; or via a statement displayed through a menu command.

tablet

Apply Quality System Regulations to Create Adequate Instructions for Use (IFU)

The Quality System Regulations, sets forth ‘Current Good Manufacturing Requirements’ for the design, manufacture, packaging, labeling, storage, installation, and servicing of all finished medical devices intended for human use.

Compliance with the requirements, if applicable, must ensure that finished medical devices are safe and effective.

Indirectly, some requirements on the Instructions for use (IFU) can be found in this part of the CFR, although often well-hidden and not summed up in one specific IFU section.

Provide device identification information in the IFU

For example, section 820.60 contains requirements on device identification. It says:

Each manufacturer shall establish and maintain procedures for identifying products during all stages of receipt, production, distribution, and installation to prevent mixups.

This could mean that when a device is installed by following the installation instructions in the IFU, the installer should be able to identify the device.

This can be accomplished by putting a picture of the device on the cover of the product, including the product name, model, type, lot or batch number, or any other information to identify the product.

cover ifu

Provide handling information in the IFU

Section 820.140 gives requirements on providing device handling information.

It mentions that manufacturers must establish and maintain procedures to ensure that ‘mixups, damage, deterioration, contamination, or other adverse effects to the product do not occur during handling’.

Providing this information in the IFU is the most common thing to hand over this information.

handling information

Provide storage information in the IFU

In section 820.150 requirements on storage information are defined.

When storage information is needed to prevent adverse effects such as contamination, damage etc., this information shall be provided.

Typically, this information is also given to the user of the medical device through the instructions for use.

storage

storage information

Provide installation information in the IFU

Section 820.170 gives requirements on the installation of devices.

When a device requires installation and/or a test procedure before it can be used, the manufacturer of the device must provide adequate installation, inspection and test procedure instructions.

Instructions and procedures must include instructions to ensure correct installation so that the device will perform as intended after installation.

These instructions and procedures must be provided with the device or made available to the person(s) installing the device in any other way.

It is the duty of the person installing the device, to make sure that the installation, inspection, and any required testing are conducted according to the provided instructions.

Basically, this means that you may want to provide some directional safety messages in your instructions for use.

installation

storage information

Inspection and any test results shall be documented by the installer to demonstrate proper installation.

You may want to add a log book in your IFU to make this easier for the installer.

logging

Make it easy to provide feedback

Section 820.180 gives requirements on keeping records. What does this mean?

It means, for example, that a manufacturer must keep a record that includes all device specifications, such as clear drawings, component specifications, software specifications, production process specifications and quality assurance procedures.

But also, packaging and labeling specifications and installation, maintenance, and servicing procedures and methods.

This is similar to the European requirement for establishing a technical file.

But also, complaints must be kept in the record. After all, the manufacturer has the duty to report certain events to the FDA.

A manufacturer must develop a procedure for receiving, reviewing and evaluating complaints. All complaints must be reviewed and evaluated to determine whether an investigation is necessary.

The instructions for use can facilitate an easy submission of complaints. There are several ways to do this.

You can include a section ‘Complaints’ (or any other meaningful name) which includes the company’s postal address, telephone number and email address.

When the information is available online as well, you may want to integrate a live chat, or submission form.

Things would be easier if information that should be collected in the record matches the information to be submitted.

This is the name of the device, the date the complaint was received, any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s) and control number(s) used, the name, address, and phone number of the complainant and the nature and details of the complaint.

complaints

Provide servicing information in the IFU

Section 820.200 of the Quality System Regulations gives requirements on providing service information.

Often, servicing is required for the safe and efficient functioning of a medical device.

It is the responsibility of the manufacturer to provide instructions on how to service the device.

Service reports must be submitted to the manufacturer, who has accordingly the duty to analyze these.

When the service report represents an event which must be reported to the FDA, the manufacturer has to label the report as being a complaint and follow the procedure as described in the previous section.

Use the Guidance on Medical Device Patient Labeling

The document Guidance on Medical Device Patient Labelling, that has been published by the FDA, serves as a supporting document and describes in more detail what content to include in the instructions.

In my opinion, this guidance document is the best information source for creating IFU for medical devices for the U.S. market.

This guidance serves two purposes.

First of all, it should assist manufacturers in the development of the instructions for use.

Secondly, it assists FDA reviewers in their review and evaluation of medical device patient labeling.

The guidance aims to make medical device patient labeling more understandable to and usable by patients (or lay persons, such as family members caring for patients).

The guidance defines medical device patient labeling as “any information associated with a device targeted to the patient or lay caregiver”.

fda guidance

Medical device patient labeling is intended to help assure that a medical device is used effectively and safely.

The following formats of medical device labeling can be distinguished: user manuals, videos, brochures, leaflets, etc.

The guidance says the following about the content to include in the Medical Device Patient Labeling.

Include risk/benefit information and/or instructions for use.

Medical device labeling typically includes risk/benefit information and/or instructions for use (IFU).

Risk/benefit information can be included when patients or lay caregivers, for example, need to provide their care practitioner with personal health information.

Based on this information, the practitioner accordingly needs to decide whether to use or not use a certain device, or make a selection between available devices for the treatment, prevention, or diagnosis of an illness.

Also, risk/benefit information helps to determine what the possible effect of a medical device on the patient may be.

risk and benefits

The development of instructions for use may be considered when patients or lay caregivers need to monitor and report on the operation or output, maintain or dispose of the device.

For example, diagnostic devices, such as a glucose monitor, mostly generate a certain output. The patient or caregiver needs to understand this output.

Also, when a device generates waste, has single use items, or contaminated reagents, these may be needed to be disposed of safely according to the instructions.

cover picture

In many cases, both risk/benefit information and instructions for use have to be provided.

For example, when the user (patient or caregiver) needs to install, operate, interpret, and manipulate a medical device.

In those cases, it is often crucial to know how to use/operate the device safely. The user needs to read and understand any warnings, precautions, and contraindications, while working with the device.

class 1 medical device

Stimulate ease of navigation: include a Table of Contents

The guidance document stipulates the importance of a table of contents (ToC) for the sake of easy navigation.

The ToC is the spine of your user manual because it helps the user to quickly and easily find the information they need and helps them to find the answer to their questions.

A clear table of contents supports the user who is navigating to the right section, without having to frantically search through the entire instruction manual.

A table of contents should be included if the instructions for use are lengthy and/or complex.

The IEC/IEEE 82079 on Information for Use mentions that when instructions for use exceed 12 pages, they shall have a table of contents.

table of contents

The standard recommends that the table of contents has enough levels to help the target audience find information efficiently, but emphasizes not to have more than four levels of headings.

heading levels

Although not mentioned in the FDA Guidance document, page numbering and an index are also useful ways to help users navigate through the instructions for use (IFU).

When you have a dynamic content delivery (e.g. instructions in the form of online help), a search bar, links to related topics, individualized delivery and context sensitive delivery, these are ways that support navigation.

search bar

Include a glossary

Consistent terminology is crucial for clear understanding of the instructions for use and thus for the safe and effective use of the device..

Medical devices often come with their own specific jargon and terminology.

Make sure you explain the meaning of terminology used, including abbreviations, in a glossary.

This is especially important when the device is intended for use by a target audience without specialist knowledge.

If a glossary is used, place it directly after the table of contents. This will alert readers that it is there to help them.

Also, add any definitions to the text, or provide links and references.

glossary

Provide descriptive information

Besides instructional information, such as operation instructions and maintenance instructions, the IFU often needs to include descriptive information.

Descriptive information is information such as a description of the purpose of the device, a general description of the device, when the device should not be used (contraindications), risks and benefits information, expectations of the device and the procedure associated with the device and general warnings and precautions.

An illustration is often the best way to describe a device, parts and its accessories. The illustration can indicate all important parts that the user needs to be aware of in the rest of the IFU, such as displays, switches, dials, and meters.

product elements

The function or purpose of each product element that contains a call out should be described in a table accompanying the illustration.

Also, include an overview of any materials in the device that may cause issues for patients with hypersensitivities.

Contraindications are conditions in which a medical device shall not be used. This is often because the risk outweighs any possible benefit.

For example, the medical device may be contraindicated for patients with epilepsy or for pregnant women.

contraindications

As discussed previously, risk/benefit information can be included when patients or lay caregivers, for example, need to provide their care practitioner with personal health information.

Make sure that risk messages always provide a source for more information, acknowledge any uncertainties (e.g. indicate when there is a lack of currently available scientific knowledge), include analogies, discuss the nature of the risk, give alternatives and discuss benefits.

Benefit messages should provide balanced benefit information. It is not sensible to rely on using marketing waffle!

Therefore, benefit messages should identify outcomes, estimate the magnitude of outcomes in a way that is meaningful for readers to understand and evaluate the benefit for individuals.

General warnings and precautions are an integral part of risk and benefit information.

benefit information

General warnings and precautions cover all specific hazard information that a user needs to be aware of before using a medical device.

You can best place the warnings and precautions at the beginning of the IFU. At least, you should act according to the guidance document….

The ANSI Z535.6 standard distinguishes directional safety instructions (such as “READ ALL INSTRUCTIONS BEFORE USE”), grouped safety messages, sectional safety messages and embedded warnings.

ansi

The first two are typically placed at the beginning. The other two respectively in the section or step to which they apply. The ANSI standard aims to warn the user of any risk that might occur.

I like that approach.

Appendix E of the Guidance document stresses this as well:

“Research shows that hazard alerts are most effective when integrated into the task information at the most relevant location.”

The appendix gives detailed additional information on definitions, purpose, content, format, location, etc related to warnings and precautions.

embedded safety messages

The appendix includes a remark about the importance of following the care regimen as explained in the IFU and how it helps to motivate the user to obey the instruction.

Regarding providing information on device expectations (and the procedure associated with the device), you may want to tell the patient what to expect before, during, and after a (surgical) procedure and/or the use of the medical device.

Consider giving instructions on post-procedural or post-operative care.

Provide operating instructions

Users need to know how to use a device. They should be given specific actionable instructions, placed in the right order.

Make clear which steps are crucial, and which ones are optional.

Besides marketing waffle, detailed technical descriptions of processes are also often unnecessary. Users care about solving their problems and finding answers to their questions: How do I run a test? How do I understand the score? How do I clean the mixer?

When writing the instructions, take into account that the user may not have medical knowledge or any experience with similar devices.

Using an eighth-grade writing style ensures that you reach most of the population.

Be clear, consistent and unambiguous!

Before giving the instruction to complete a task, start with stating the purpose and expected outcome of a task.

purpose test

purpose

Give clear setup instructions

Let the user know if the setup of the device needs to be done by a qualified person. For example, by adding the following warning:

WARNING! Set up shall be carried out by a properly qualified and skilled installer.

WARNING! Installation and commissioning shall be carried out by a properly qualified and skilled installer with due observance of the applicable rules and legislation.

If the lay user is allowed to set up the device, include the following information (if relevant):

Provide checkout instructions

Sometimes, a device requires checkout procedures for the sake of safety and effectiveness.

For example, this may be required directly after installation, to make sure that the device is properly installed.

Or after maintenance, to make sure that the device is ready for use again.

The below example is taken from a GE Healthcare manual and shows an example of checkout after installation.

checkout procedure

Typically, a checkout procedure includes:

Give operating instructions

Operating instructions are some of the most crucial parts of the instructions for use. Make sure that these instructions are clear and easy-to-follow.

The operating instructions may include any information on special preparation the user needs before operation. Does he or she need to wash their hands, warm up the device or should the temperature remain between certain values?

warning

Add any warnings or safety instructions specifically related to operation and again: place them immediately before the corresponding task or instruction.

If the user can recognize an incorrect operation, tell him/her how to recognize this.

embedded warning

As always, the operating steps should be numbered and placed in a logical sequence.

Tell the user what results to expect when an instruction has been executed. For example:

example ifu

expected results

Save some space in the IFU to add any user-specific instructions.

Provide a telephone number of the person to call if there is a problem.

emergency

Give instructions on how to monitor the device

For some devices it may be important to carefully monitor the correct functioning of the device.

If this is important to your specific medical device, emphasize this importance in the IFU.

Instruct the user how he/she can make sure that the device acts as intended.

Provide clear examples of what to check exactly and to make sure that the medical device still works properly.

For example, this is a very specific instruction given in order to monitor if a certain message appears on the GUI:

monitor

Give accurate cleaning instructions

As cleaning and sterilization of a medical device is often of huge importance for its safe use, cleaning instructions should be drafted with great care.

Always make sure to list all the supplies needed in a section specifically intended for step-by-step cleaning procedures.

cleaning instructions

Tell the user the recommended frequency of cleaning the device, what the cleaning accomplishes (e.g. reducing hazards) and what the results of failure to clean would be.

Include warnings and precautions for cleaning agents, explain what the consequences are of using inadequate cleaning solutions or methods and add instructions required for the correct disposal of the suggested cleaning agents.

disposing cleaning material

Give maintenance instructions

Maintenance instructions give the user all necessary directions and information to maintain the device effectively.

They should clearly describe what maintenance actions are the responsibility of the user and what the frequency of those actions should be.

If certain maintenance instructions should be performed by qualified personnel only, this shall be indicated in the IFU.

maintenance

Consider creating separate booklets if several users are involved for performing maintenance tasks, for example one booklet for the end-user of the device and one for the maintenance engineer.

Give storage instructions

Sometimes medical devices need to be stored for shorter or longer periods. The storage conditions may affect the device.

Provide storage instructions to prevent any malfunctioning or device damage.

Start with indicating to the user what the reason for proper storage is and what possible consequences are of improper storage.

Explain how the device needs to be properly prepared for storage and define the storage conditions.

When longer periods of storage affect the device, for example when there is a risk of leaking batteries, inform the user about these risks.

If appropriate, include information on extended storage in relevant sections, such as the installation/setup, checkout, operation, or maintenance section.

storage information

Give failure time information

Most devices have a limited lifetime. But even if they do not, it is always good to mention the technical lifespan in the IFU. Why?

Because this defines how long the device can be operated safely when the instructions in the IFU are being followed.

It limits the amount of information that has to be provided in the IFU.

For example, when the lifespan is 1 year, you won’t need to give any maintenance information on parts that need to be replaced after that year.

Defining the lifespan limits your responsibility, warranty and your liability.

You can also add information on the importance of following the instructions and what the results and expectations are when the device fails, or the instructions are not followed.

In case of failure, let the user know if it is dangerous (or safe) when the device fails and what the effect might be on the patient.

This information should also be part of the Risks/Benefits section.

Give disposal instructions

Not only waste, cleaning cloths, and single-use items, but also the device itself, should be disposed of in a proper way.

When appropriate, instruct the user how to safely dispose of the device.

disposal information

Provide information on accessories and other devices

Medical devices often come with accessories, add-ons, additional items, extra features etc.

Use the IFU to list all appropriate content areas for each accessory. You can do this by including a separate section to describe any accessories, or by adding the information to the section to which it applies.

If relevant, put a general warning at the beginning of the instructions for use, advising users of problems that may occur if they use accessories other than those you recommend.

I often choose to add this information in the intended use section.

accessory information

Give troubleshooting information

A proper troubleshooting section is crucial to help users solve any problems.

Typically, this is a separate section on the instructions for use.

The best way to present troubleshooting information, is by giving error information (what happens?), the cause (what is the problem?), and the solution (how do I solve the error?) in a 3-column table.

error information

Note: The principles of Minimalism have another interesting way of providing troubleshooting information, called ‘supporting error recognition and recovery’, which is the 3rd Minimalism principle.

According to this principle, error recognition information and error solving information are provided within the instructions/steps, so not in a separate chapter, as in a troubleshooting chapter.

The idea behind this is that you want to help people to manage errors when they occur.

This means that you have to support detection and you have to support correction. These are the two minimalist aspects of error management.

The following example shows how error recognition and solving are embedded in the task domain.

error recognition

Provide any additional information

Lastly, the guidance document for FDA reviewers recommends to include all kinds of additional information, such as clinical study information, disease and self-care information, adverse events, user assistance information, warranty information and travel or international use information.

It also mentions an index, which we discussed previously, and the date of printing.

index

Instruction for Use (IFU) Checklist

That is a lot of information that we have discussed. Therefore this checklist can be of use to make sure you include all information in the IFU:

Apply FDA Approved Medical Device Standards

Most CFR requirements contain label requirements and regulate essential, but general, device information such as expiration date, intended use, dose quantity etc.

We have not found specific requirements on the IFU yet. Why?

Because it is hard to give all the requirements for each of the wide variety of devices that are available.

The contents of the IFU of a pacemaker are different from that of an MRI scan. The user of each device requires different information in order to use the device safely and effectively.

Standards are being developed to define requirements for specific medical devices or groups of medical devices.

For example, there are standards for wheelchairs and ambulatory electrocardiographic systems.

The same applies to dental cartridge syringes and steam sterilizers.

The FDA approves certain standards when they think the standards meet all safety and effectiveness goals of the FFDCA.

These can be national approved standards, such as the ANSI/ADA 48-2-2010 (R2015) LED Curing Lights.

ansiada

Or international standards, such as ISO 10939:2017 for slit-lamp microscopes.

iso

Approved standards can be found in the Federal Register.

Consult relevant approved standards, as they often include information on the IFU.

Federal Register FDA

For example, the FDA approved standard IEC 61010-1 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements contain requirements on the documentation.

Apply Labeling Requirements for Electronic Products According to 21 CFR Part 100

Now we’ve discussed the labeling requirements on all kinds of medical products, let’s have a closer look at electrical medical devices.

That’s because a medical device can also be an electronic device.

Actually, almost all medical products that we at INSTRKTIV deal with are electronic medical devices.

After all, electronics means that there are more features and complexity, so the more complex a medical device, the more instructions for use should come with the device.

See these two types of electronic medical devices:

amabrush

mri scan

Mark the device with a tag declaring compliance with applicable standards

A medical device must bear a tag or label containing certification and compliance information.

This tag or label shall be permanently fixed or inscribed.

It should have a prominent position, be legible, and easy to see when the product is ready for use.

Possibly, a mandatory standard may have its own requirements on the certification. In these cases, the standard shall be followed.

When one or more standards apply to an electronic device, the manufacturer must make sure that the dealer receives a certification that the device meets the applicable standard(s).

Provide the name and address of the manufacturer

The label of an electrical medical device must contain the name and address of the manufacturer. Abbreviations, such as “Co.” and "Inc." may be used.

When a medical device is sold under a different name other than the manufacturer, the full name and address of the entity under whose name the device is sold should be used, but only when this person or company has made the manufacturer’s name known to the Center for Devices and Radiological Health.

Also, this tag shall be permanently affixed to the device and be clearly visible and legible.

Provide the place, month and year of manufacture

As we’ve just seen that the date format of a medical article should have the form yy-mm-dd, for electronic devices, the place, month and year of manufacture should be mentioned.

The place of manufacture may take the form of a code, but only when this code has been made known to the Center for Devices and Radiological Health.

The month and year must be in a clear and legible format. The year must be shown as a four-digit number:

Manufactured: December 2020

Apply Labeling Requirements for Investigational Devices

When a medical device is either NOT approved for marketing in the U.S., or IS approved for marketing, but is being clinically evaluated for a new indication, a device is considered investigational.

Investigational devices are regulated and an Investigational Device Exemption (IDE) must be obtained from the FDA.

An IDE allows an investigational device to be used in order to collect safety and effectiveness data required to support a premarket approval (PMA) application or a premarket notification submission to the FDA.

All investigational devices must contain a label indicating the name and place of business of the manufacturer/packer/distributor, the quantity of contents (if appropriate), and the following statement:

CAUTION - Investigational device. Limited by Federal (or United States) law to investigational use.

label investigational device

The label or other labeling (such as the instruction for use) must give all relevant hazards, warnings, precautions, contraindications, adverse effects and interfering substances or devices.

Statements that are false or misleading in any particular way are strictly forbidden.

Among such statements are messages that the device is safe or effective for the purposes for which it is being investigated.

Investigational devices for laboratory animals must bear the following statement:

CAUTION - Device for investigational use in laboratory animals or other tests that do not involve human subjects.

Apply U.S. Labeling Requirements for In Vitro Diagnostic Products

In vitro diagnostic devices form a special group of medical devices and have their own set of requirements on the labeling and instructions for use (IFU).

In vitro diagnostic medical devices are tests used on biological samples to determine the status of a person's health.

According to the website of the FDA, in vitro diagnostic products are “those reagents, instruments, and systems intended for use in diagnosis of disease or other conditions, including a determination of the state of health, in order to cure, mitigate, treat, or prevent disease or its sequelae. Such products are intended for use in the collection, preparation, and examination of specimens taken from the human body.”

in vitro diagnostic

Section 809 of the Code of Federal Regulations (CFR) Title 21 applies to in vitro diagnostic products and Part 809.10 gives requirements on the labelling of in-vitro diagnostic products.

Requirements on the instructions for use (IFU) fall under the subsection ‘Labeling Requirements for Inserts and Outer Packaging’.

According to the CFR requirements, all required information must be in one place and in the format and order as specified in the requirements.

I have included this list below.

An exception is made when information is not applicable, or when a standard is available that requires differently.

Labels of reagents that are intended as a replacement in a diagnostic system, may be limited to contain only the information necessary to adequately identify the reagent and to describe its use in the system.

Multiple purpose instruments used for diagnostic purposes (and not committed to specific diagnostic procedures or systems) can be labeled with just the points annotated by an asterisk (*).

quantative test